CLEAN ROOM EQUIPMENT

Bio Safety Cabinet

Biosafety Cabinet Manufacturers in India

A biosafety cabinet is Ventilated laboratory workspace for safely working with materials contaminated with (or potentially contaminated with) pathogens requiring a defined biosafety level. Airflow is drawn around the operator into the front grille of the cabinet, which provides personnel protection. Because cabinet air has passed through the exhaust HEPA filter, it is contaminant-free (environmental protection). The purpose is to maintain the sterility of materials kept inside the chamber and process the chemicals.

What is a BSL?

Biological Safety Level (BSL) is a biocontainment designation developed by the American Biological Safety Association (ABSA) to protect personnel from exposure to harmful, potentially pathogenic microbes in a research or manufacturing environment.

 

    What differentiates BSL designations from one another?

    In the United States, the Center for Disease Control specified four general Biosafety Levels, adopted from the ABSA designations. Each increasing BSL corresponds to greater risk, building upon the previous level to provide greater protection.

    • BSL-1: for work with well-characterized microbes that do not present a serious health risk, such as laboratory strains of E. coli. This designation does not require engineering controls (i.e. a biosafety cabinet) and can be carried out on the bench with minimal personal protective equipment.
    • BSL-2: for work with moderately hazardous microbes known to cause infectious diseases in humans. An example of this would be Staphylococcus aureus. This level requires the use of a biological safety cabinet with negative-pressure containment, and exhausted air must be filtered prior to release.
    • BSL-3: for work with potentially lethal microbes that are easily aerosolized, such Mycobacterium tuberculosis. This level requires restricted access to the work environment, immunization of personnel (where available), and the use of respirators. Work must be performed in a primary engineering control such as a biosafety cabinet, and air must be filtered before it can be exhausted.
    • BSL-4: for work with high-risk, airborne pathogens that frequently cause lethal infections, such as Ebola. Requires all work to be carried out in a Class III BSC. In addition, personnel must use full-body, positive-pressure cleanroom suits. All materials must be decontaminated prior to leaving the BSC.
     

    Which Bio Safety Cabinet is appropriate to which BSL?

    Biological safety cabinets were designed to provide localized protection from potentially biohazardous substances. In general, the BSL level corresponds to BSC class, indicating the BSC type most appropriate for each BSL application. The National Science Foundation designates three classes for BSCs:

    •  Class I: a negative-pressure hood in which air is drawn from the room into the enclosure and subsequently exhausted after treatment with UV-C germicidal light and HEPA filtration. These enclosures protect personnel, but do not protect products within the hood.
    • Class II: protect both operator and product. These units provide a HEPA-filtered laminar flow of air to protect the product from contamination, much like a laminar flow hood. Air is then directed into negative pressure plenums where it can be re-filtered or exhausted into a ducting system. Four common types of Class II BSCs are distinguished by air velocity as well as the percentage of air that recycled versus exhausted.
    • Type A1 (formerly Type A): minimum inflow velocity of 75 ft/min. 70% of air is recirculated; 30% is ducted. The contaminated plenum is surrounded by the lab environment.
    • Type A2 (formerly Type A/B3): minimum inflow velocity of 100 ft/min., with a negative plenum surrounding the contaminated plenum. 70% of air is recirculated; 30% is ducted.
    • Types B1 and B2 have dedicated hard-ducted exhaust lines, with 70% air exhaust in Type B1 and 100% air exhaust in Type B2.
    • Class III: provide the highest level of protection when working with high-risk pathogens. These enclosures are gloveboxes with airtight seals. Operators use glove ports to manipulate materials within the enclosure, never coming into direct contact with any materials. Exhaust air is HEPA filtered and directed into a ducting system.
     

    BSCs, BSLs, and ISO Classes

    When it comes to working with hazardous or potentially infectious material, protecting personnel must be the primary concern. However, offering personnel protection does not have to come at the cost of risking product integrity. Class II BSCs provide protection to both personnel and product simultaneously. They provide an ISO class 5 clean environment for product manipulation while ensuring that exhaust air is filtered and ducted out of the environment. Thus, a class II BSC is the ideal solution when working with moderately hazardous material and both personnel and product safety are required

    Protect your research and your team’s safety with Brinda Pharma’s Biosafety Cabinets – the leading Bio Safety Cabinet manufacturer of high-quality, durable and reliable Biosafety Cabinets in India. Contact us today to learn more and ensure the safety of your lab!

    Bio Safety Cabinet Features, Accessories and Applications

    • Internal area of SS-316
    • The main structure of SS-304/ GIPC
    • Grill for HEPA – SS-304
    • Worktop – perforated work (SS-316/304)
    • Dynamically and statically balanced motor and blower
    • HEPA Filters: Efficiency of 99.999% (H14) down to 0.3 mic.
    • Motor blower assemblies, tubular motor
    • LED light, UV light On/Off Panel , Auto Option available
    • Wrist Support
    • Auto Sash Opening with foot-operated switch
    • Complete S.S 316 Option
    • Glove Port option for increased operator & product safety
    • Cell Processing,
    • Bio Processing,
    • Cell Culture
    • Bio Safety Cabinet for Laboratory 
    • Research & Development
    • Virus and Vaccine Production
    Previous slide
    Next slide
    ModelClassInternal Dimensions
    (W X D X H) mm
    External Dimensions
    (W X D X H) mm
    Construction
    BSC-I-1270-650-SClass I1270 X 650 X 6601420 X 750 X 2160External G.I.
    Powder Coated

    Internal S.S. 316
    BSC-I-1870-650-SClass I1870 X 650 X 660 2020 X 750 X 2160External G.I.
    Powder Coated

    Internal S.S. 316
    BSC-II-A1-1270-650-SClass II-A11270 X 650 X 6601420 X 750 X 2160External G.I.
    Powder Coated

    Internal S.S. 316
    BSC-II-A1-1870-650-SClass II-A11870 X 650 X 660 2020 X 750 X 2160External G.I.
    Powder Coated

    Internal S.S. 316
    BSC-II-A2-1270-650-SClass II-A21270 X 650 X 6601420 X 750 X 2160External G.I.
    Powder Coated

    Internal S.S. 316
    BSC-II-A2-1870-650-SClass II-A21870 X 650 X 660 2020 X 750 X 2160External G.I.
    Powder Coated

    Internal S.S. 316
    BSC-II-B1-1270-650-SClass II-B11270 X 650 X 6601420 X 750 X 2160External G.I.
    Powder Coated

    Internal S.S. 316
    BSC-II-B1-1870-650-SClass II-B11870 X 650 X 660 2020 X 750 X 2160External G.I.
    Powder Coated

    Internal S.S. 316
    BSC-II-B2-1270-650-SClass II-B21270 X 650 X 6601420 X 750 X 2160External G.I.
    Powder Coated

    Internal S.S. 316
    BSC-II-B2-1870-650-SClass II-B21870 X 650 X 660 2020 X 750 X 2160External G.I.
    Powder Coated

    Internal S.S. 316
    BSC-III-1270-650-SClass III1270 X 650 X 6601420 X 750 X 2160External G.I.
    Powder Coated

    Internal S.S. 316
    BSC-III-1870-650-SClass III1870 X 650 X 660 2020 X 750 X 2160External G.I.
    Powder Coated

    Internal S.S. 316

    We at Brinda Pharma are committed to fulfilling the needs of our clients; that’s why we do 100% customization based on your needs.

    For complete Customization, We follow the standard practice of understanding the URS – User Requirement Specification. Even if you don’t have URS, our experienced Sales Team can guide you to make one or choose one from the Industry Standards.

    All Ceiling Suspended Laminar Air Flow comes with the following:

    • DQ (Design Qualification)
    • FAT (Factory Acceptance Test)
    • Quality & Test Certificates For All Components (Traceability Certificate)
    • IQ (Installation Qualification)
    • OQ (Operation Qualification)
    • Maintenance Manual
    • Spare Part List
    • Warranty Certificate

      Contact Us Today, For a Free Quote

      * For information about how we handle personal data, see our privacy statement.