Cleanroom Rules
Cleanroom Rules Leave personal items outside the controlled environment If items such as keys, wallets, phones, etc. must be brought into
It works on a very basic principle. First ambient room air is sucked through a preliminary filter to capture larger particles to extend the lifespan of the secondary HEPA filter. The air is then blowed towards the HEPA Filter through Fans. Later HEPA Filter generates a smooth stream of clean air projected vertically across the desired workspace. The downward airflow from the ceiling laminar airflow unit clears out and weakens all impurities, contamination and creates a workspace free of particles for improved sterile operations and processes.
The entire system is mounted/hanged from the ceiling like a fan, Thus this particular system is know as Ceiling Suspended Laminar Air Flow System.
Accessory | Description |
---|---|
Clean Down Timer | Timer feature for scheduling and automating the cleaning process, ensuring consistent cleanliness. |
Panel/Wall Mounted Control Box | A seprate control box mounted on panel/wall for accessibility. |
CG Screen/Distributor | A Textile fabrix to imporve air flow and air velocity. |
PVC Curtain | To improve the process and minimize the external particles a PVC Curtain is fixed vertically on the sides of the System upto desired length. |
Is it possible to install a Ceiling Suspended Laminar Air Flow in existing facility?
Yes It is. Our team has designed the system so that it can be installed in an existing facility. We do also customize the system according t0 the requirements.
How can I choose the right Ceiling Suspended Air Flow system for my facility?
When choosing the system, consider factors such as the size of your area/room where it is required. One of the key things is to make sure that the system is able to cover the particular process/machine under it. We also have to make sure there is sufficient room height is available so it can easily installed in the facility. Our expert sales professionals can guide you to choose the right size and configurations for your application.
Where are ceiling suspended laminar airflow systems typically used?
These systems are commonly used in industries such as pharmaceuticals, biotechnology, healthcare, electronics manufacturing, and research laboratories. They are essential in areas requiring strict contamination control, such as sterile production, clean rooms.
What are the benefits of using a ceiling suspended laminar airflow system?
CSLAF used for minimizing airborne contamination by creating the clean air environment which improve product quality by reducing the risk of contamination during production activity. Also enhance the operator safety by minimizing the exposure to hazardous substances or pathogens.
How do you maintain a ceiling suspended laminar airflow system?
Regular maintenance is crucial to ensure the system's optimal performance. Upon installation a periodic filter replacement, cleaning as well as cleaning of diffusers and grill. A periodic calibration of pressure gauges is required for accuracy.
Does Brinda Pharma Technologies provide delivery & installation to my destination?
As a leading manufacturer and exporter we provide delivery orders across the world. We also provide products with installation & validation, as well as an annual maintenance contract to ease your operation. Reach out to our Sales Team to discuss any questions related to Delivery & Installation.
Cleanroom Rules Leave personal items outside the controlled environment If items such as keys, wallets, phones, etc. must be brought into
A clean room is a room where the concentration of particles is minimized. Temperature, humidity and pressure parameters are also
Accessory | Description |
---|---|
Pressure Gauge | Accurate pressure gauge for monitoring and maintaining optimal airflow and pressure levels. |
Clean Down Timer | Timer feature for scheduling and automating the cleaning process, ensuring consistent cleanliness. |
Pilot Light | Indicator light to signal the operational status of the system, providing visual feedback to users. |
Buzzers | Audible alarms or buzzers to alert users of any anomalies or safety concerns, enhancing system safety. |
UV Light | UV light for sterilization and improved contamination control |
HMI+PLC Based Control System | To monitor and change various parameters from a single touchscreen interface which makes entire system easy to operate. |
Flame Proof System (FLP) | A Complete or partial Flame Proof System along with FLP Lights to accommodate safe working environment and prevent fire hazards. |